Class I Biosafety Cabinet MBC-DI Series
The principle of the product
Fresh air ventilation cabinet without external piping, top fan and filter. Turn on the switch, so that the formation of negative pressure fan operation in the cabinet, to ensure no chemical spill.
The main wind direction from top to bottom, so that the cabinet air through the filter, clean air is exhausted from an air outlet from the top. On the one hand to clean the cabinet volatile
chemicals, on the other hand, the laboratory air into the cabinet, also can purify the air in the laboratory. Simple and convenient installation, energy saving and environmental protection,
no consumption of air conditioning, air purification in the laboratory, protect the health of employees.
Safety production: the basic principle of the cabinet
1.Weighing safety cabinets designed for precise weighing and negative pressure air flow design, to ensure a good control effect and filtering effect. To ensure the maximum experimental safe.
2.Ensure the maximum experimental safe.
Security cabinet: weighing technology
1.Ensure the seismic effect, the real psychological board, the weighing accuracy of 10
2.Ensure the constant surface wind speed, large LCD screen display surface wind speed, the wind speed is too high or too low alarm.
3.Fan speed can be adjusted, the HEPA filter with the time increasing, the pressure loss will be larger, in order to ensure the normal to the surface wind speed CPU monitoring of surface wind
speed is too high or too low self regulating, and achieve the security value.
Specification:
Model | MBC-800DI | MBC-1000DI | MBC-1300DI | MBC-1600DI | MBC-1800DI |
Internal Size (W*D*H)mm | 798*560*760 | 998*560*760 | 1298*690*760 | 1598*690*760 | 1598*690*760 |
External Size (W*D*H)mm | 800*620*1146 | 1000*620*1146 | 1300*790*1146 | 1600*790*1146 | 1600*790*1146 |
Filter | 1Module, 2pcs | 1Module, 2pcs | 2Modules, 4pcs | 3Modules, 6pcs | 4Modules, 8pcs |
Fan | 1pc | 1pc | 2pcs | 3pcs | 4pcs |
Air Flow | 240m³/ h | 240m³/ h | 480m³/ h | 720m³/ h | 960m³/ h |
Air Velocity | 0.4-0.6m / s |
Noise | 41dba |
LED Lamp | 5W*1 |
Consumption | 30W | 30W | 60W | 90W | 120W |
Display | LED display:the temperature in the cabinet; surface wind speed; fan speed; alarm setting; Optional: 1. Air quality changes in laboratory and on-line monitoring PPM value; 2. Real-time monitoring of the filter outlet concentration of chemicals and displays the PPM value |
Control System | Microprocessor |
Material | Front panel: 8mm thick acrylic sheet |
Epoxy resin worktop |
Plate: 6mm thick PMMA embedded 1.2mm galvanized steel sheet |
Bearing Part: 2.0mm thick galvanized steel, strong bearing |
Main Body: Cold-rolled 1.2 steel with anti-bacteria epoxy polyester coating. |
The filter module: polypropylene (PP material) |
Interface | Anti-electromagnetic interference power line. |
Power Supply | 110/220V±10%, 60/50Hz |
Optional Accessory | Filter: Filter for organic gases or Filter for the acid gas and organic gas; H14 HEPA filter |
Base: Fixed table (with shelves) or Fixed frame work; |