JMD Series Level Pulse Vacuum Sterilizer/Autoclave
Feature:
1. Mobile door—automatic elevator doors, automatic sealing with air pressure drive seal
2. The two—door channels, with safety interlock and liquid program with temperature safety door interlock.
3. Nice chamber surface
Surface smoothness, Ra≤0.5; No welding part or bolt in the chamber which effectively protect the items to be sterilized. Steam baffle plates are surrounded by silicon rubber protecting the chamber surface. Design use life of 1500 cycles.
4.The Structure and Material:
1) Main structure for euro rectangle ring stiffener structure, patented main body sealing structure design
2) Material:External and internal: 304 stainless steel; Jacket: Q245R pressure vessel material.
3) The acid-proof stainless steel chamber and jacket are automatic plasma robot-welded to guarantee long time and trouble-free operation.
5. Touchable LCD display of sterilization status, temperature, pressure, time, jacket and chamber temperature, pressure curve.
6. Communication interface with the quality traceability system in disinfection supply center, with free software communication protocol.
7. Sterilization programs for fabric, instrument,lumen devices, implantable devices, rubber, liquid, gravity, quick, drying, customer procedures, manual control, B&D test, PCD test and leakage test.
8.Built-in miniature stylus printer to print real-time process data of the temperature, pressure, sterilizing time, operator.
9.Intelligent maintenance system with remote monitor alarm of the door seal replacement, air filter clean-up & replacement, door elevating lubrication and maintenance.
10.Independent steam admission design, avoids jacket carbon polluting the sterilization items and avoid jacket pressure decrease affecting drying effect.
11.Piping system: 304 stainless steel, auto welder welding; High quality imported valves: US ASCO Angle seat valve, Japan SMC solenoid valve, pneumatic valve, German SIHI straight association-type water ring vacuum pump
12.Water-saving, noise reduction system, reducing the vacuum pump circulating water consumption by 30%.
13.Auto water drain device: auto detection of jacket,chamber temperature and pressure, auto draining, ensures the steam saturation.
14.Auto control of drainage water temperature: ≤ 55℃.
15.Air-filter: antibacterial rate of 99.97% at 0.22um.
Specification:
Technical Parameter
Interior design pressure | -0.1~0.3Mpa |
Design Pressure of Jacket | 0.3Mpa |
Built-in steam generator | 0.8Mpa |
design temperature | 150℃ |
maximum working temperature | 139℃ |
maximum working pressure | 0.25Mpa |
opening pressure of safety valve (inner chamber) | 0.28Mpa |
opening pressure of safety valve (jacket) | 0.28Mpa |
opening pressure of safety valve (evaporator) | 0.7Mpa |
Pressure of internal chamber hydraulic pressure test | 0.4Mpa |
jacket test pressure | 0.5Mpa |
Temperature display accuracy | 0.1℃ |
Pressure display accuracy | 0.1Kpa |
Temperature control accuracy of sterilization | ±0.2℃ |
temperature uniformity | ±0.5℃ |
Fluctuating vacuum frequency | 1~99 |
Pulsating online | 80Kpa |
Pulsating offline | -90Kpa |
final vacuum | -97Kpa |
Positive pressure pulse frequency | 0~9 |
Vacuum leak rate | ≤0.13Kpa/min |
Specification Model
Model | Inner dimension | Volume (L) | Outline dimension (L×W×H) | Weight (kg) | water Kg/C | Steam Kg/C | Power Supply | Power |
W×H(mm) | Depth(mm) |
JMD-0.3 | 550×550 | 1000 | 300 | 1388×950×1800 | 750 | 150 | 19 | 380V,50HZ | 12KW |
JMD-0.3D | 1000 | 300 | 1388×950×1800 | 750 | 150 | 19 | 380V,50HZ | 12KW,30KW |
JMD-0.45 | 672×672 | 1000 | 450 | 1388×975×1980 | 950 | 200 | 22 | 380V,50HZ | 12KW |
JMD-0.45D | 1000 | 450 | 1388×975×1980 | 1035 | 200 | 22 | 380V,50HZ | 12KW,30KW |
JMD-0.675 | 1500 | 675 | 1888×975×1980 | 1120 | 340 | 36 | 380V,50HZ | 2KW |
JMD-0.675D | 1500 | 675 | 1888×975×1980 | 1200 | 340 | 36 | 380V,50HZ | 2KW,50KW |
JMD-0.81 | 1800 | 810 | 2188×975×1980 | 1450 | 400 | 47 | 380V,50HZ | 3KW |
JMD-0.81D | 1800 | 810 | 2188×975×1980 | 1550 | 400 | 47 | 380V,50HZ | 3KW,54KW |
JMD-0.91 | 672×910 | 1500 | 910 | 1860×1325×2100 | 1500 | 450 | 47 | 380V,50HZ | 3KW |
JMD-0.91D | 1500 | 910 | 1860×1325×2100 | 1600 | 450 | 47 | 380V,50HZ | 3KW,54KW |
JMD-1.1 | 1800 | 1100 | 2160×1325×2100 | 1850 | 500 | 55 | 380V,50HZ | 35KW |
JMD-1.1D | 1800 | 1100 | 2160×1325×2100 | 1950 | 500 | 55 | 380V,50HZ | 35KW,60KW |
JMD-1.23 | 672×1230 | 1500 | 1230 | 1884×1365×2232 | 1900 | 600 | 65 | 380V,50HZ | 35KW |
JMD-1.52 | 1850 | 1520 | 2234×1365×2232 | 2050 | 750 | 80 | 380V,50HZ | 35KW |
JMD-2.15 | 972×1230 | 1800 | 2150 | 2234×1670×2232 | 2550 | 1000 | 106 | 380V,50HZ | 45KW |